Device Recall IMRIS ORT300, Removable Operating Room Table 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Deerfield Imaging, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79488
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1266-2018
  • 사례 시작날짜
    2018-02-23
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Table, operating-room, ac-powered - Product Code FQO
  • 원인
    The operating room table may drift while in use, which could potentially result in unintended movement of the table.
  • 조치
    The firm initiated their field notification on 02/23/2018 by letter. The notice stated the following: "It has recently come to our attention that the ORTT2X/100/200/300 tables may drift while in use, which could potentially result in unintended movement of the table. This may pose a risk of injury to the patient. To minimize these risks, IMRIS is taking the following corrective actions: " ORT T2X/100/200 tables: The tables will be tested for drift symptoms. No further action will occur for tables that pass the drift test. For tables that do not pass the drift test, a hydraulc fluid flush and filter change will be performed. " ORT300 tables: all customer sites with these tables shall have the hydraulic fluid flushed and exchanged for new and the filter element will be changed.A drift test will also be performed. IMRIS has identified that your facility has one or more of these OR Tables which require evaluation. Please confirm details on page 2 and return your response to IMRIS. IMRIS Customer Service will contact you to implement this corrective action. IMRIS recommends these preventive actions: " Per the ORT Operator Manuals, the table should be tested daily in all standard positions and inspected for anomalies. Before each surgery, inspect the table for any visibly loose components or hydraulic fluid leaks and check that the headrest and tabletop safety latches are secured. .Please contact customer support via our service line at 1-866-475-0525 or Email: customersupport@imris.com if you have any questions or concerns."

Device

  • 모델명 / 제조번호(시리얼번호)
    (a) 114093-000, Serial Numbers: 10005695, 10002998, 10005456, 10007377, 10002997, 10005612, 10004372, 10002841, 10002255, 10005613 (b) 114093-600, Serial Numbers: 10007411, 10002996, 10004515
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    US, Canada, China, Australia, Sweden, Germany
  • 제품 설명
    IMRIS ORT300, Removable Operating Room Table, Part Numbers: || (a) 114093-000 || (b) 114093-600
  • Manufacturer

Manufacturer