Device Recall Infant Radiant Warmer 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 GE Healthcare, LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77287
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2323-2017
  • 사례 시작날짜
    2017-05-05
  • 사례 출판 날짜
    2017-05-19
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2018-07-12
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Warmer, infant radiant - Product Code FMT
  • 원인
    Ge healthcare has recently become aware of a potential safety issue related to loose screws in the heater head of the infant warmer system (iws). hot screws from the heater head of the iws could fall onto the bed if the heater head assembly has been improperly serviced. this situation can be clinically hazardous because thermal injury to a patient could result. two injuries have been reported as a result of this issue.
  • 조치
    GE Healthcare sent an Urgent Medical Device Correction letter dated May 5, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Product Correction : Attached to this letter, we provide instructions as part of a Service Manual Addendum on how to correct the issue. Please add this new Addendum to the Service Manual of your device(s) and train the affected users accordingly. Please acknowledge that you have received this letter and that you understand that an action needs to be taken on your part to correct this issue by filling out and returning the attached Customer Response form. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.

Device

  • 모델명 / 제조번호(시리얼번호)
    Medical device listing number: D103053, D103054, D103056, D103057, D183529
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA (nationwide) Distribution.
  • 제품 설명
    Infant Warmer System (IWS)
  • Manufacturer

Manufacturer

  • 제조사 주소
    GE Healthcare, LLC, 3000 N Grandview Blvd, Waukesha WI 53188-1615
  • 제조사 모회사 (2017)
  • Source
    USFDA