U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Fluorometer, for clinical use - Product Code KHO
원인
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
조치
Tecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk.
US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC
제품 설명
Infinite 200 PRO in vitro diagnostic. || Product Usage: The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384 well microplates conforming to the ANS/SBS standards.