Device Recall Innova SelfExpanding Stent System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Boston Scientific Corporation 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    58910
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-2502-2011
  • 사례 시작날짜
    2011-05-19
  • 사례 출판 날짜
    2011-06-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2014-02-04
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Stent, superficial femoral artery - Product Code NIP
  • 원인
    Boston scientific corporation is conducting a recall of the innova self-expanding stent system because they have received 6 complaints involving no deployment/partial deployment of the innova stent. potential health hazard events resulting from this type of failure include increased procedure time, vessel wall injury, and/or emergency surgery to remove the partially deployed stent. this recall d.
  • 조치
    Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter, dated May 13, 2011, to all affected customers via registered mail. The letter explained the issue, identified affected product, stated that further distribution and or use of the devices should cease immediately, and asked that devices in inventory to be segregated and immediately returned to Boston Scientific. An Account Reply Verification Tracking Form was also asked to be returned. Boston Scientific Sales Representative were assist in the return process by providing a Returned Goods Authorization (RGA) number. Customers were instructed to contact their local Sales Representative for any questions.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lots: 14166856, 14190182, 14218448, 14222766, 14267441, 14267975, 14271561, 14300373..
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Worldwide Distribution - USA (nationwide) and the countries of Product removal in US is being managed through SuperNOVA Clinical Trial suspension in accordance with IDE #G100291. OUS distribution to include: Austria, Finland, France, Germany, Italy, Netherlands, Spain, and Switzerland.
  • 제품 설명
    Boston Scientific Innova Over-the-Wire Self-Expanding Stent System, Catalog No. 39180-08607. 8 mm x 60 mm x 75 cm (stent diameter x stent length x delivery system length). Made in USA, Two Scimed Place, Maple Grove, MN 55311 USA. The Innova Self-Expanding Stent System is comprised of two components: the implantable endoprosthesis and the stent delivery system. The stent is a laser cut self-expanding stent composed of a nickel titanium alloy (Nitinol). On both the proximal and distal ends of the stent, radiopaque markers made of tantalum increase visibility of the stent to aid in placement. The stent is constrained within a 6F (2.1 mm maximum OD) delivery system. The delivery system is a triaxial design with an outer shaft to stabilize the stent delivery system, a middle shaft to protect and constrain the stent, and an inner shaft to provide a guidewire lumen. The delivery system is compatible with 0.035 in (0.89 mm) guidewires. When ready to be implanted, the stent is deployed by retracting the middle shaft of the delivery system. A radiopaque marker at the distal end of the delivery system aids in visibility during deployment. As the stent is exposed to body temperature, it expands to appose the vessel wall. || The Innova Self-Expanding Stent is available in a variety of diameters and lengths. || The delivery system is also offered in two working lengths (75 cm and 130 cm). || The Innova Self-Expanding Stent System is indicated for the treatment of peripheral vascular lesions
  • Manufacturer

Manufacturer

  • 제조사 주소
    Boston Scientific Corporation, 1 Scimed Pl, Maple Grove MN 55311-1565
  • 제조사 모회사 (2017)
  • 제조사 의견
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source