Device Recall Konica Minolta ImagePilot software 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Konica Minolta Medical Imaging USA, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    80220
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2282-2018
  • 사례 시작날짜
    2018-05-09
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, image processing, radiological - Product Code LLZ
  • 원인
    Panasonic has issued a mandatory product safety notice, indicating a possibility of battery ignition (fire) in cf-c2 "toughbook" laptops. since konica minolta uses this laptop in some of its product configurations, it notified its customers to utilize a bios utility for the laptops to minimize the risk of an incident while panasonic develops a final solution.
  • 조치
    The firm issued an Important Product Recall Information letter to their affected customers dated May 3, 2018, delivered via UPS Overnight Delivery. The letter notified customers of the Panasonic recall and how it affected the Konica Minolta product. The letter instructed customers to utilize a BIOS Utility for CF-C2 laptops to minimize the risk of an incident while Panasonic develops a final solution. Customers were provided with Technical Bulletin which detailed the steps. Correction Option 1: Update BIOS Environmental Setting using Utility Tool 1. Download the BIOS Utility Tool from the Panasonic Website: ftp://ftp.panasonic.com/ computer/ software/ Panasonic-CF-C2-Charge-Tool.zip 2. Load the Utility Tool onto the CF-C2 Desktop 3. Double-click to run batch file: ChgCtrlUty.bat 4. Optional tool: If Integrated Portable, an external keyboard may be required. Visit KM University at https://km-university.com/ for video instructions: https://km-university.com/course/notification-of-mandatory-safety-bulletin-of-panasonic-toughbook-cf-c2/ Correction Option 2: Manual change to the BIOS Environmental Setting (for Integrated Portable, an external keyboard will be required) 1. If power is on, turn the power off and restart prior to following instructions. 2. Upon boot-up, at Panasonic splash screen, continuously hit the F2 key on the keyboard to access the BIOS. a. Go to the "MAIN" tab associated with the BIOS setting. b. Select "ENVIRONMENT Current Status" setting. c. Ensure "HIGH TEMPERATURE" is selected. Otherwise change this setting to high and click ENTER. d. Go to the "EXlr' tab associated with the BIOS setting. e. Select "SAVE VALUES AND REBOOr' and hit ENTER key on the keyboard. f. A prompt will appear to state, "SAVE VALUES". Select YES and hit ENTER. If you have any questions, need a keyboard, require assistance with these instructions, and/or a safety verification please contact our Customer Care Team at (800) 945-045

Device

  • 모델명 / 제조번호(시리얼번호)
    All version numbers.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.
  • 제품 설명
    ImagePilot, Model No. D9MA
  • Manufacturer

Manufacturer

  • 제조사 주소
    Konica Minolta Medical Imaging USA, Inc., 411 Newark Pompton Tpke, Wayne NJ 07470-6657
  • Source
    USFDA