Device Recall LATITUDE Patient Management System Communicator 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Boston Scientific CRM Corp 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    59421
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-3183-2011
  • 사례 시작날짜
    2011-05-25
  • 사례 출판 날짜
    2011-09-09
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-12-04
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Implantable cardioverter defibrillator (non-crt) - Product Code LWS
  • 원인
    Boston scientific has identified two occurrences where the latitude system did not process a data payload, which is the information uploaded to the latitude system from a remote interrogation of the implanted device. these occurrences resulted in non-displayed payloads because the data is not posted to the website. if the payload is not displayed, the pertinent data will not be available on the.
  • 조치
    Boston Scientific notified all affected customers by telephone on May 25, 2011. The customers were informed that there was an interrogation for one of their patients that was not able to be displayed on the website. Boston Scientific detected that a payload failure had occurred, but could not conclude what information was contained within the payload. Each patient has had previous and subsequent payloads that have processed and displayed on the LATITUDE website.

Device

  • 모델명 / 제조번호(시리얼번호)
    SN 035074 SN 032915
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Nationwide Distribution-in the states of Florida and Texas.
  • 제품 설명
    LATITUDE¿ Patient Management System Communicator, Model 6482. || The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Boston Scientific pulse generator and transfer data to a central database.