Device Recall LeadCare Plus Blood Lead Testing System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Magellan Diagnostics, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    76991
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-1926-2017
  • 사례 시작날짜
    2014-11-29
  • 사례 출판 날짜
    2017-05-18
  • 사례 상황
    Completed
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Lead, atomic absorption - Product Code DOF
  • 원인
    Leadcare plus and leadcare ultra blood lead testing systems may generate falsely low lead concentrations in some blood samples. the issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the becton dickinson edta vacutainer tubes, a collection device used with the testing system.
  • 조치
    Magellan issued a follow-on Recall notification related to the FDA Safety Notification dated May 17, 2017. This current recall follows FDAs recommendation to remove venous blood samples as an allowable sample type from all LeadCare products. This recall will supersede and replace the prior Recall notifications. Notifications issued by : UPS Letter (US), Registered Letter, E-mail, or Phone calls form May 23-May 25. Questions: Reba Daoust, Director of QA/RA 978-248-4811 or rdaoust@magellandx.com Refer to Firm's website: http://www.leadcare2.com/getmedia/2fd8a90f-5c64-4058-b6ab-5db13512f0fe/Letter-to-our-customers-5-17-17.pdf.aspx. The notification of May 2017 supercedes the firms previous communications:. Magellan Diagnostics, notified customers by letter dated Nov 29, 2014 and provided with the problem, mitigation action( customers have been advised to implement a minimum 24-hour incubation of the blood-treatment reagent mixture prior to analysis) . Customers were instructed to return a completed FaxBack form via fax to 978-600-1480. A revised letter issued May 1, 2017 recommending that customers make the following workflow changes: 1. Discontinue collecting blood samples for lead testing in Becton Dickinson (BD) EDTA Vacutainer¿ tubes: Lavender and Tan tops. 2. Discontinue the 24 hour incubation for sample preparation. On 5/17/17 FDA issued News Release - FDA warns Americans about risk of inaccurate results from certain lead tests- manufactured by Magellan Diagnostics may provide inaccurate results for some children and adults. Blood drawn from a vein may provide results that are lower than the actual level in the blood. FDA is warning laboratories not to use Magellan Diagnostic lead tests wit blood drawn from the vein. At this time all Leadcare systems can be used with blood drawn from a finger or heel stick. Based on FDA/s warning the firm issued notification letter to users on 5/xx/17.