Device Recall LIFEPAK 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Medtronic Emergency Response Systems, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    31111
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-0562-05
  • 사례 시작날짜
    2005-02-03
  • 사례 출판 날짜
    2005-03-10
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2006-08-21
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Automated External Defibrillators (Non-Wearable) - Product Code MKJ
  • 원인
    Potential to not detect patients at the low end of the patient impedance range.
  • 조치
    On 2/3/05 the firm issued a 'Medical Device Correction' letter dated January 2005 to their customers. The letter advised customers of the issue and stated a representative from the firm would visit each facility to update affected unit. On 2/25/05 a nationwide press release was issued. On 3/1/05 the firm began telephoning each customer, this was followed by another letter titled MEDICAL DEVICE RECALL. The letter advises of the issue and recommends defibrillators remain in service when no alternative is available until the unit can be corrected. Each unit will be replaced or corrected. On 4/22/05 the firm issued another letter, via certified mail return receipt, expanding their recall. Letter was issued to all customers affected by the expansion of the recall. On 4/22/05 the firm issued a press release advising of the expansion of the recall. The firm's web site also provides a listing of affected units.

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial numbers - various, noncontiguous Part numbers - 3005400-XXX, D3005400-XXX, U3005400-XXX
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Devices were distributed to 750 domestic consignees dn 16 international consignees.
  • 제품 설명
    LIFEPAK 500 AED.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Medtronic Emergency Response Systems, Inc., 11811 Willows Rd Ne, Redmond WA 98052-2003
  • Source
    USFDA