U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
System, Test, Blood Glucose, Over The Counter - Product Code NBW
원인
Meter labeling (owner's booklet and quick reference guide) is incorrect as it indicates that the meter displays the message 'do you need a snack' for glucose values between 20 and 70 mg/dl; whereas the onetouch ultrasmart meter displays the message, 'do you need a snack?' for glucose values from 20 through 59 mg/dl.
조치
On 6/9/05, the firm initiated the recall and their notification was via letters informing its customers of the corrective action.
All One Touch UltraSmart Meters shipped with OneTouch UltraSmart Owner''s Booklet AW 060-788-01A are affected by this Field Correction. Part Number: Description: 020-524-01, System,OUS,US; 020-524-01F, System,OUS,US; 020-524-01Y, UltraSmart Meter; 020-524-02, System,OUS,EAS-C,US; 020-524-03, System,OUS,EAS-S,US; 020-524-04F, OT UltraSmart System Kit V. 4 USA; 020-524-04Y, UltraSmart System; 020-524-05, UltraSmart System -EAS-C; 020-524-06, UltraSmart System -EAS-S; 020-524-10, OT UltraSmart System Kit; 020-524-10F, OT UltraSmart System Kit; OT UltraSmart System Kit V. 10 USA; 020-524-10Y, OT UltraSmart System Kit; 020-524-11, System, EAS-C, OUS, US; 020-524-12, System, EAS-S, OUS, US; 020-524-13, OT UltraSmart System; 020-759-02F, OT UltraSmart Meter V.2 USA; 020-759-03F, OT UltraSmart Meter V.3 USA; 020-761-01F, Starter,OUS,US; 020-761-02, OT UltraSmart Starter; 020-761-02F, OT UltraSmart Starter; 020-761-04, OT UltraSmart Starter; 020-761-04F OT UltraSmart Starter; 020-761-05 OT UltraSmart Starter Kit; 020-761-05F, OT UltraSmart Starter Kit; 020-926-01, Dist Kit,SM,OTU,US; 020-930-01, Dist,Kit,OTU/OUS,US; 020-971-01, Disp,Counter,Ultra Kits; 021-075-01, UltraSmart SVP; 021-113-01, Display, Hanging, OUT/OUS, US; 021-133-01, Display, Counter, OTU/OUS Kits; 021-135-01, Span Display, Hanging, OTU, US; and 021-166-01, Display, Hanging, OTU/OUS Kits.
Product was released for distribution to 40,763 consignees worldwide. The firm provided the following distribution information: 124 wholesaler dealers, 224 distributors, 140 retailers and 40,275 consumers/users involved in the recall. The recall product was distributed in all States within the U.S. In addition, OneTouch UltraSmart meter kits with the affected UltraSmart Owner''s Booklets (AW 060-788-01A) have been distributed internationally in Canada, France, Germany, Italy, United Kingdom, Benelux, Sweden, Finland, Norway, Denmark, Spain, Portugal, Switzerland and Austria. The recall was appropriately extended to the consumer/user level; i.e., the wholesalers/distributors, pharmacists, and patients who received the recalled products.
제품 설명
LifeScan brand One Touch UltraSmart Meter Owners Booklet AW 060-788-01A; || All One Touch UltraSmart Meters shipped with OneTouch UltraSmart Owner''s Booklet AW 060-788-01A are affected by this Field Correction.