Device Recall LifeStent FlexStar Stent System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Bard Peripheral Vascular Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    49405
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0069-2009
  • 사례 시작날짜
    2008-08-29
  • 사례 출판 날짜
    2008-10-10
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-04-06
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Catheter - Product Code FGE
  • 원인
    Some lifestent flexstar systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible.
  • 조치
    The customer letters for the tip to sheath gap were sent on 8/29/2008 with return receipt. The letter advises that C.R. Bard has determined that some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible. Per the letter, the firm's analysis has determined that a gap between the tip of the delivery system and the delivery system sheath may be associated with difficult advancement to the treatment site. Users are instructed to inspect all devices to verify that there is not a gap between the tip and sheath (orange colored guidewire lumen is visible), prior to use and to not use any devices that exhibit a tip to sheath gap condition. If devices that exhibit a tip to sheath gap are identified, customers are advised to call Customer Service at 1-800-321-4254 to return the device and a replacement will be provided at no cost. The letter also advises that the Instructions for Use for the LifeStent FlexStar System will be updated to include the inspection step which is provided in the letter.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Numbers: FR6J0264, and  FR6K0311
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    Yes
  • 유통
    Nationwide
  • 제품 설명
    Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards Lifesciences LLC Irvine, CA 92614, Model EX090403CD. The LifeStent FlexStar self-expanding biliary stent system is intended for use in the palliation of malignant strictures (neoplasms) in the biliary tree.
  • Manufacturer

Manufacturer