Device Recall Maxi Move Patient Lift with Scale 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Arjo, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    37064
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0796-2007
  • 사례 시작날짜
    2007-01-05
  • 사례 출판 날짜
    2007-05-11
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2008-09-02
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Patient Lift - Product Code FSG
  • 원인
    The hanger bar assembly may detach from the maxi move lifters with scale, due to an incorrect torque setting on the two bolts holding the scale in place.
  • 조치
    Arjo sent Urgent Device Field Correction Recall Notice letters dated 1/5/07 to the end user accounts who received the affected lifts, advising them of the potential for the hanger bar assembly to detach. They were informed that an Arjo Field Service Technician would call them within five business days to schedule the time for him to visit and perform the inspection and correction to the lifts involved. If they did not hear from the technician within that time, they were instructed to contact the Arjo Quality Department at 800-323-1245, ext. 6140. The accounts were also requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model number KMMB4ELU2FUS, serial numbers SEE0624509, SEE0624510, SEE0624511, SEE0624512, SEE0624513, SEE0624514, SEE0624515   (Manufactured week 24 of 2006 through week 28 of 2006. You can check the model and serial number of your unit by looking at the serial number label on the mast of the lift, located underneath the removable battery. Serial numbers are all structured SEEYYWWXXX, where ¿ YY = the last two digits of the year,  ¿ WW = the two digits of the production week, and ¿ XXX = the sequential number the unit was in production that week ¿ Example: SEE06024001 = the first unit produced in week 24 of 2006.   If your equipment has one of the model numbers listed above and a serial number beginning with SEE0624 through SEE0628, it is affected by this notification, regardless of the last three digits of the serial number.)
  • 유통
    Nationwide
  • 제품 설명
    Maxi Move Patient Lift with Scale; Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden; Model KMBB4ELU2FUS
  • Manufacturer

Manufacturer

  • 제조사 주소
    Arjo, Inc., 50 Gary Ave Ste A, Roselle IL 60172-1684
  • Source
    USFDA