Device Recall Medline Strider Maxi 4 Scooter 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Medline Industries Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    36072
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0037-2007
  • 사례 시작날짜
    2006-07-20
  • 사례 출판 날짜
    2006-10-17
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2009-09-02
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    scooter - Product Code INI
  • 원인
    Under conditions of heavy use, signs of overheating of the main battery cables have been observed.
  • 조치
    Medline sent 'Voluntary Field Correction' letters dated 7/20/06 to the direct accounts on the same date, informing them that the Strider Maxi 3 and Maxi 4 scooters have main battery cables that may overheat under conditions of heavy use, and that Medline will provide an upgrade kit to replace the battery cables. Any questions were directed to 800-289-9793, press 1, then press 3 and ask to transfer to 4674 (8:00 AM to 5:00 PM CST). Follow-up letters were sent to the distributors and end-users on 10/31/06. The end users were informed of the problem with the scooters, and were requested to call Medline at 800-289-9793, press 1, then press 3 and ask to transfer to 4674 (8:00 AM to 5:00 PM CST) to have the scooter upgraded. The end users were requested to inform the service technician that hey have one of the affected Maxi Scooters, and provide him with their full name, phone number and street address, as well as the name, address and/or phone number of teh store where the Medline Scooter was purchased. Medline will arrange with the dealership or retail store to upgrade the scooter at no cost to the end user.

Device

  • 모델명 / 제조번호(시리얼번호)
    model MDS807650 - red and model MDS807650B - blue, all serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide
  • 제품 설명
    Medline Strider Maxi 4 Scooter; a four wheeled battery operated scooter; manufactured by KYMCO, Kaohsiung, Taiwan for Medline Industries, Inc., One Medline Place, Mundelein, IL 60060 USA; model MDS807650 - red and MDS807650B - blue
  • Manufacturer

Manufacturer

  • 제조사 주소
    Medline Industries Inc, 1 Medline Pl, Mundelein IL 60060-4486
  • Source
    USFDA