U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Aseptic Transfer Set - Product Code GAZ
원인
The mentor aseptic transfer set contains a component, the bd 60ml luer-lok syringe, which is under recall by becton dickinson due to a package integrity issue.
조치
Firm sent notification, along with a copy of the Becton Dickinson recall letter, to consignees on 08/18/08. Consignees asked to inspect inventory of Mentor Aseptic Transfer Sets for affected BD 60mL syringes and to complete and return the included Device Accountability Worksheet. For additional information, please contact Mentor Customer Service at 800-235-5731.
Aseptic Transfer Sets: All lots. Syringes: Reorder Number: 309653; Lot numbers beginning with the numbers 5, 6, or 7 and also the following lot numbers. 8010954, 8014571, 8031115, 8031117, 8031119, 8031124, 8031133, 8038154, 8038158, 8045564, 8045567, 8058890, 8058896, 8058898, 8058900, 8058902, 8072401, 8072409, 8072414, 8072419, 8079645, and 8079656.
Worldwide Distribution; product sent to all states (except Alaska), Washington DC and Puerto Rico. Product also sent to the following countries: Australia, Canada, Costa Rica, Korea, New Zealand and South Africa.
제품 설명
Mentor Aseptic Transfer Set, Cat. #350-8400 (contains BD 60mL Luer-Lok Syringe). Aseptic transfer set manufactured by Mentor Texas, LP, Irving, TX 75038 (BD 60 mL Luer-Lok syringe manufactured by Becton Dickinson, Franklin Lakes, NJ 07417).