Device Recall MHITM2000/VERO 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    72991
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1244-2016
  • 사례 시작날짜
    2015-12-15
  • 사례 출판 날짜
    2016-03-24
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-04-20
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Accelerator, linear, medical - Product Code IYE
  • 원인
    Due to operator console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "completed (or discontinued)" to "untreated" under specific conditions.
  • 조치
    Due to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" if all the following conditions are met: (1) At least one patient has been treated on the day; (2) A major error occurs when VeroTM/MHI-TM2000 is operated in a mode, other than Clinical Mode, which means Daily Check Mode or Maintenance Mode. (3) The major error is NOT associated with a function that uses therapeutic/ MV radiation parameters, such as an error in Isocenter Check function, etc. Note 1: If the major error is associated with a function that uses therapeutic/MV radiation parameters, the problem that treatment completion status is changed will never occur. The Operator Console changes the treatment completion status in the course of processing the major error. Note 2: Although treatment completion status is changed to "Untreated", its delivered dose result has been correctly saved into R&V; System. It has not been lost. Advice on Actions by Customer/Users: 1. Please make sure to perform the operations as described below in case a major error occurred in a mode other than Clinical Mode. (1) Cancel the error. (2) Click [Clinical] button on the left of Operator Console screen. (3) Verify, on R&V; System screen, that treatment completion status of the patients who were treated on the day are all correct. The treatment completion status is indicated in different colors as follows depending on the status: Untreated: BLUE, Completed: GRAY, Discontinued: RED. Customers requiring support concerning this issue, please contact the Customer Service Representative: 1-800-597-5911 (Mon - Fri) or via email us.support@brainlab.com. If any other problem is detected, please stop operation and contact the Customer Service representative.

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial numbers: 201901, 201902, 201903, 201904, 201905, 202901, 202902, 202903, 202904, 202905, 202906, 203901, 203902, 203903, 203904, 203905, 203906, 203918, 203919, 203921, 203922, 203924, 203925, 203927 and 203929.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Distributed in the states of Florida, New York, Ohio & Texas, and the countries of France, Germany, Japan, Italy, Korea, & Belgium.
  • 제품 설명
    MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) || Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.