Device Recall MiniInfuser Model 150XL Infusion Pump 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Baxter Healthcare Corp. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    62881
  • 사례 위험등급
    Class 3
  • 사례 연번
    Z-2268-2012
  • 사례 시작날짜
    2011-03-04
  • 사례 출판 날짜
    2012-08-24
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-08-27
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Pump, infusion - Product Code FRN
  • 원인
    Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the mini-infuser will not operate. the reason being that the mini-infuser battery spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. the iec 60086-2 and ansi 08 standards specify a minimum.
  • 조치
    Baxter sent a Safety Alert letter on January 11, 2011 for the InfusO.R. Pump. Baxter expanded the scope of the recall to include the Mini-Infusers System and sent an Urgent Device Correction letter dated March 4, 2011, via First Class mail to all affected customers. The letters informed the customers that the Mini-Infusers use the same size battery and incorporate the same battery contact interface as the InfusOR pump, and that the same feature that inhibits electrical contact in over-the counter Energizer C size batteries may also be incorporated by other battery manufacturers. The customers were requested to verify that the InfusOR and Mini-Infuser will power-on before use; if the device does not power on, replace the batteries with a different brand of size C batteries and verify that the device will power-on before use; and to place a copy of the letter with each operator's manual for the InfusOR and Mini-Infuser System for future reference. For Technical questions call Baxter Global Technical Services at 1-800-843-7867. For question regarding this communication contact the Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.

Device

  • 모델명 / 제조번호(시리얼번호)
    product codes 2M8170 and 2M8170R, all serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    45,865 units
  • 제품 설명
    Mini-Infuser Model 150XL Single Speed Infusion Pump, an Rx battery operated syringe infusion pump; || Product Usage: These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician. Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073