U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Calibrators, drug specific - Product Code DLJ
원인
Complaint from the sole distributor that the third party control material is recovering high, and outside the published range.
조치
Microgenics sent a Product Removal notification letter dated June 15, 2011 to all affected customers. The letter identified the affected product, the problem, and the necessary actions to be taken. Customers were instructed to discontinue use and dispose of the affected product per local waste ordinances, review all patient results reported with the use of this lot, and retain a copy of this notification for their records. The letter states that if the affected product has been forwarded to another laboratory, a copy of the letter is to be provided to them. An alternative calibrator is available to customers impacted. It can be ordered by calling Thermo Fischer Scientific Customer Service at 1-800-232-3342.
Questions regarding the information should be forwarded to Abbott Customer Service at 1-877-4ABBOTT.
Nationwide Distribution (USA) in the state of ILLINOIS
제품 설명
Multigent Vancomycin Calibrators; || Manufactured by Microgenics Corp, Fremont, CA. || Distributed by: Abbott Diagnostics, Abbott Park, IL || Product Usage: || The Multigent Vancomycin calibrators are intended for use in the calibration of the multiagent vancomycin || Vancomycin is a commonly used antibiotic to treat staphylococcus and other gram-positive bacterial infections.