Device Recall NeuViz 64 Multislice CT Scanner System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Neusoft Medical System Co 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    71306
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1693-2015
  • 사례 시작날짜
    2014-03-08
  • 사례 출판 날짜
    2015-06-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-12-05
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, x-ray, tomography, computed - Product Code JAK
  • 원인
    The a-plane compensator is made of teflon which will become fragile due to x-ray exposure passing through it. the thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. failure is indicated by pin or ring artifact in the image.
  • 조치
    Neusoft planned action: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Neusoft Medical Systems Co., Ltd will implement actions to monitor the exposures of the tube and when the exposure number is close to 300,000 seconds, the A-plane compensator will be replaced. The Food and Drug Administration (FDA) district office coordinator should be notified when the last A-plane compensator is replaced. 3. The Customer Information Letter has been sent to customers as a corrective action. The maximum exposure number of 5 US systems is 191830 seconds as of Apr.20, 2015, so the firm will monitor the data continuously and replace the A-plane compensators before the defined threshold. 4. Neusoft will replace all compensators that have exceeded the 300,000 seconds free of charge and will replace all compensators that fail during warranty and the compensators failed less than 300,000 seconds though out of warranty free of charge. You may proceed or continue with implementation of the CAP (subject to the conditions noted above) For further questions please call (281) 453-1205

Device

  • 모델명 / 제조번호(시리얼번호)
    NeuViz 16/64 Series
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US Distribution including Puerto Rico and to the states of : IL, MA, TX and SC., and to the countries of : China, Malaysia, Lebanon, Ecuador, Egypt, United Arab Emirates,Kazashstan, Tunisia, Angola, Brazil, Uzbekistan, Lebanon, Cote D' ivoire, South Africa, Montenegro and Tajikistan.
  • 제품 설명
    Neusoft NeuViz 64 Multi-slice CT Scanner System