Device Recall NEXES VEN IVIEW DAB DET KT JPNUS EXPORT 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Ventana Medical Systems Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78942
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0740-2018
  • 사례 시작날짜
    2017-12-19
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    in vitro diagnostic - Product Code NJT
  • 원인
    Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (hrp) dispensers (part of the iview, ultraview and optiview detection kits, as well as cintec plus cytology kit) and with hematoxylin ii.
  • 조치
    On Dec 19th all Roche affiliates (the direct consignees) worldwide were notified via Safety Board Notification (SBN) and subsequent Field Safety Notification (EX-US) and UMDC TD00090 (US) SBN and FSN revised Dec 20th due to 3 affected part numbers missed on original letter. The instructions are as follows: "Actions taken by Roche Diagnostics All affected product has been placed on hold. Ventana has reworked all product in its inventory and is in the process of manufacturing new lots for distribution and replacement of customer affected kits. Customers will be notified when corrected product is available. Actions to be taken by the customer/user Affected kits may continue to be used by customers until corrected product is available, however Ventana is mandating that the affected IHC detection kits (iView, ultraView, OptiView) must only be used in conjunction with same-slide controls. These controls must be appropriate for each assay and capable of detecting false negative results due to a complete or partial reagent dispense failure. CINtec PLUS Cytology does not have the capacity for same slide controls, so system-level controls should be maintained. ultraView SISH Detection is used for HER2 analysis, and employs internal positive controls; no external control is required. For assays that directly relate to clinical therapy decision making (e.g. ER/PR, HER2, ALK, etc.), it is additionally important to select a same slide positive control tissue with sufficient sensitivity to detect small decreases in intensity that may cause borderline positive cases to appear as negative (e.g. HER2 2+ vs. 1+). Although the use of same slide controls is considered optimal laboratory practice and strongly recommended by Ventana, customers may revert to standard run controls once non-impacted product is received. In order to reduce the risk of this issue impacting patient care, customers not using same slide controls as a standard practice should follow their

Device

  • 모델명 / 제조번호(시리얼번호)
    UDI 04015630970254, Lot Number: Y15392
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
  • 제품 설명
    NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI 05266084001 in vitro diagnostic
  • Manufacturer

Manufacturer

  • 제조사 주소
    Ventana Medical Systems Inc, 1910 E Innovation Park Dr, Oro Valley AZ 85755-1962
  • 제조사 모회사 (2017)
  • Source
    USFDA