Device Recall Oncentra External Beam Treatment Planning Software 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Elekta, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    72869
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0604-2016
  • 사례 시작날짜
    2015-12-18
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System,planning,radiation therapy treatment - Product Code MUJ
  • 원인
    When using the option "tumor overlap fraction" in vmat planning it has been observed that in rare cases the system does include an organ at risk as target volume. this could result in open mlc, and open jaws in areas away from the target volume.
  • 조치
    Elekta sent an Important Field Safety Notice 806-01-ETP-001 to all affected customers on December 18, 2015. The notice informs users of the issue and that should the issue occur it could result in overdosage of organs at risk outside the intended treatment volume. Users are strongly advised to perform proper Quality Assurance for all treatment plans before delivery of the first fraction to the patient. To prevent the issue from occurring, the workaround is to not use the Tumor Overlap Fraction for VMAT planning.and not to use the Tumor Overlap Fraction for VMAT planning. The issue will be solved in Oncentra External Beam version 4.5.2. The notice included an acknowledgement form which is to be returned to Elekta. For questions contact your local Elekta representative.

Device

  • 모델명 / 제조번호(시리얼번호)
    Versions 3.3 and higher
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.
  • 제품 설명
    Oncentra External Beam - VMAT - Radiation therapy planning system || Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Elekta, Inc., 400 Perimeter Center Ter NE Ste 50, Atlanta GA 30346-1227
  • 제조사 모회사 (2017)
  • Source
    USFDA