Device Recall OPT Croutons synthetic bone substitute 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Acumed LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    75904
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1019-2017
  • 사례 시작날짜
    2016-12-05
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-03-22
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Filler, bone void, calcium compound - Product Code MQV
  • 원인
    Opt croutons, part number 65-0210-s, is recalled due to the result of inadequate documentation of the sterilization dose used in production. product sterilized at the site did not receive the minimum dose required per the new performance qualification.
  • 조치
    Acumed sent the Urgent Recall Notice letter, dated December 1, 2016, to US customers on December 5, 2016. Customers are advised to follow the instructions provided by Acumed: IMMEDIATELY stop using and/or distributing the product and complete the following actions: 1. Perform a physical count of your inventory. 2. Remove the affected OPT Croutons, such that the croutons are out of service. 3. Record this data on the survey found following this recall letter. 4. Fill out the survey and return it via email to RecaiiNotification@acumed.net. 5. For any product that needs to be returned: Domestic Customers: please contact Acumed Agent Inventory via email at Agentlnventory@acumed.net or contact your agent directly to be assigned a Return Material Authorization (RMA) number and applicable shipping information. International Customers: please contact Acumed Business Services via email at BusinessServices@acumed.net or via phone at (877) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada) to be assigned a Return Material Authorization (RMA) number and applicable shipping information. 6. Once completed, please return the OPT Croutons that have been removed to Acumed using the Return Label provided by Acumed and included in your replacement shipment. Please contact Acumed Business Services via email at BusinessServices@acumed.net or via phone at (877) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada) with any questions regarding this notification.

Device

  • 모델명 / 제조번호(시리얼번호)
    lot 549/12.001. Part number 65-0210-S.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    Yes
  • 유통
    distributed in FL, IN, KS, MI, MO, NE, OH, PA, TX and Peurto Rico.
  • 제품 설명
    OPT Croutons is a synthetic bone substitute designed for the repair of bony defects and is made of pure beta-tricalcium phosphate. The tricalcium phosphate is a soluble calcium phosphate salt.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Acumed LLC, 5885 NW Cornelius Pass Rd, Hillsboro OR 97124-9432
  • 제조사 모회사 (2017)
  • Source