U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Nidek inc. received information from our manufacturer nidek co. japan that the shown axial
lengths were much shorter at al-scan measurement.
조치
Dear Doctor/Recall letters were sent to affected customers on July 7, 2016. The letters describe the issue and provide instruction should the issue occur. The letters inform users that Nidek will reach out in the next few weeks to schedule installation of new software that will solve the problem.
Questions or concerns may be directed to:
Customer Service 1.800.722.0219; Monday- Friday 8 am to 5pm PST
E-mail: AL-Scan_support@noritsuservice.com.
US Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.
제품 설명
OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. || Ophthalmic: The OPTICAL BIOMETER AL-Scan is a measuring device that can measure as a single unit the values necessary to calculate the power of an IOL for cataract surgery such as axial length, corneal curvature radius, and anterior chamber depth. It measures these necessary values successively through a non-contact optical measurement method.