U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
원인
Orthoscan inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. specifically, the system software allows a user to activate the digital zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user.
조치
Orthoscan sent an Urgent Recall Notice, dated August 30, 2017, to customers, informing them of the issue and how to address it. Customers are to use the attached list of serial numbers to determine if their device is affected. Customers may continue to use their affected device with caution until the on-site correction has been made. Users should avoid non-conforming workflows by taking the measures outlined in the letter. The firm will contact customers to arrange implementation of the corrective action at no cost to the customer. If there are further questions or concerns, please call National Customer Service Team at (866) 949-4346.
Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.
제품 설명
Mobile Mini C-arm system Part # 1000-0001. || Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.