Incorrect parameters: if the user switches from one surgeon profile to another surgeon profile after completing anatomical landmark acquisition tasks, the application will continue to use the initial surgeon profile parameter values in the calculation of the predicted hip-knee-ankle alignment estimates, which may have different resection or component thickness parameters than are appropriate for.
조치
Zimmer notified consignees via letter dated 8/19/08, which advised them not to change the selected surgeon profile after the acquisition of landmarks has begun, and the issued will be fixed with the next software release.
All units with software versions/distribution numbers Orthosoft-unicondylar 1.0.2.31, Orthosoft-unicondylar 1.0.1.28 and Orthosoft-unicondylar 1.0.0.0.111.