Device Recall Pentax Inc 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Pentax of America Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    75249
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0757-2017
  • 사례 시작날짜
    2016-05-31
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Bronchoscope (flexible or rigid) - Product Code EOQ
  • 원인
    The eb-1970uk ultrasound video bronchoscope shipped between the clearance date of april 2014 and september 2015 contained the operation and reprocessing ifus that were in use prior to the 2014 clearance, and therefore did not include the cleared operation and reprocessing ifus.
  • 조치
    Pentax sent a US Urgent Field Correction letter to all affected customers with a field correction/field safety response form via USPS on May 31, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Revised instructions were as follows: The instructions have been revised in the following manner: ¿ Operation IFU o Integrated all models of each family into the IFU o Modified the indications for use language to be consistent with similar products o Enhanced the explanation of the operation of the device o Revised and added cautions and warnings ¿ Reprocessing IFU o Included more specificity regarding reprocessing activities to aid in the proper method for reprocessing the endoscopes, including: ¿ number of brush and rinse steps, ¿ quantity of fluid volumes, ¿ use of additional text, figures, cautions, and warnings Identification of Affected Endoscope Models Table 1 provides an overview of the Affected Endoscope models, and the part number and revision of the historical and revised IFUs. The accompanying USB flash drive contains all the revised IFUs identified below. Please distribute only those IFUs which support the products owned by your facility. Table 1 Affected Models Historical Version of IFUs Revised IFUs EB-1970UK Ultrasound Video Bronchoscope Z870 Owners IFU S016_R02 Operator IFU S021_R04 Reprocessing IFU No.388_R01 Addendum No.392_R00 Addendum EB-70K series Video Bronchoscopes Z705 Operation IFU S015_R04 Operator IFU S020_R04 Reprocessing IFU No.388_R01 Addendum No.392_R00 Addendum Z706 Reprocessing IFU EB-75K series Video Bronchoscopes S003 Owners IFU Customers were instructed to ensure that user departments discard the prior versions of the Instructions for Use and ensure that all users read and carefully follow the revised Operation and Reprocessing IFUs to ensure the proper use and cleaning of the affected PENTAX endoscopes. Customers were advised to contact their PEN

Device

  • 모델명 / 제조번호(시리얼번호)
    EB-1970UK
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    Ultrasound Video Bronchoscope || The EB-1970UK, Ultrasound Video Bronchoscope, is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Pulmonary Track including but not restricted to organs, tissues, and subsystem: Nasal Passage, Pharynx, Larynx, Trachea, Bronchial Tree (including access beyond the stem), and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Pentax of America Inc, 3 Paragon Dr, Montvale NJ 07645-1782
  • 제조사 모회사 (2017)
  • Source
    USFDA