Device Recall RayStation Therapy Treatment Planning System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 RAYSEARCH LABORATORIES AB 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    65136
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1419-2013
  • 사례 시작날짜
    2013-02-15
  • 사례 출판 날짜
    2013-05-28
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-02-17
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System,planning,radiation therapy treatment - Product Code MUJ
  • 원인
    Under some circumstances, imported ct, mr and pet images are offset from their true positions by one pixel in one or two directions. users must be aware of this information to avoid incorrect dose calculations during treatment planning.
  • 조치
    RaySearch Laboratories sent an Urgent Field Safety Notice, Medical Device Correction letter dated January 28, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to educate their planning staff and all users about the recommended workarounds contained in the letter. Customers were also asked to inspect their product and identify all installed units that were infected and then complete the attached Reply Form and e-mail or scan a copy of the completed form to the firm. Customers were instructed to pass along this information to all those who need to be aware of this Medical Device Correction Customers with questions were instructed to contact the RaySearch support email at support@raysearchlabs. For questions regarding this recall call +46 8 545 06130.

Device

  • 모델명 / 제조번호(시리얼번호)
    Software version and build numbers 2.0.0.15, 2.5.1.89, and 3.0.0.251 (to determine if the version used is affected, bring up the About box in the RayStation application and check the build number reported0.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA including Connecticut, Florida, Massachusetts, Missouri, New York, Ohio, and foreign distribution to(Canada, Italy, Australia, Germany, Netherlands, and Switzerland.
  • 제품 설명
    RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. || RayStation is a software system designed for treatment planning and analysis of radiation therapy.
  • Manufacturer

Manufacturer