Device Recall RayStationTherapy Treatment Planning System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 RAYSEARCH LABORATORIES AB 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    73575
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1712-2016
  • 사례 시작날짜
    2016-02-10
  • 사례 출판 날짜
    2016-05-25
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-04-24
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System,planning,radiation therapy treatment - Product Code MUJ
  • 원인
    A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. the views can become unsynchronized with the stored data if simultaneously right clicking, pressing ctrl-s or ctrl-z while the left mouse button is held down. this bug does not affect dose computations, which are based on the stored system state.
  • 조치
    On February 10, 2016, RaySearch Laboratories distributed Field Safety, Medical Device Correction notices and Reply Forms to their customers via email. To determine if the software version you are using is affected, open the About RayStation dialog in the RayStation application and check if the build number reported. This bug does not affect dose computations. Dose is always computed based on the stored system state. Actions to be taken by the user include the following: Always release the left mouse button before using other keys or mouse buttons, If the warning message "Temporary items should not be visible on approval" is displayed or if there is some suspicion that the bug may have triggered in structure definition, either review the structures in the Plan approval dialog, or close and reload the patient case and review the structures again. All customers should educate planning staff and all users about this workaround. All customers are advised to inspect product and identify all installed units with the referenced software version number, then confirm that you have read and understood the notice by completing and returning the reply form. The completed reply form should be returned to Freddie Cardel, freddie.cardel@raysearchlabs.com, fax 888-501-7195. Customers with questions can call 1-877-778-3849, available east coast business hours all US working days.

Device

  • 모델명 / 제조번호(시리얼번호)
    3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13 or 4.7.4.4
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    **Domestic** AR, AZ, CA, CT, CO, DE, FL, GA, IL, LA, MI, MD, MS, MA, NY, NV, NJ, OH, OR, PA, NC, RI, TN, WA & Washington D.C; **Internationally** Algeria, Australia, Austria, Belgium, Canada, China, France, French Polynesia, Germany, Iran, Italy, Japan, Netherlands, New Zealand, Norway, Republic of Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom
  • 제품 설명
    RayStation Therapy Treatment Planning System Stand-alone Software 3.0, 3.5, 4.0, 4.5 and 4.7., designed for treatment planning and analysis of radiation therapy.