Anesthesia Delivery Unit - Product Code BSZ
원인
There exists the possibility that the adu fresh gas control unit (a-fgc1) can shut off during normal handling of the anesthesia machine and result in no anesthetic agent output.
조치
An Urgent Medical Device Correction letter dated March 21, 2006, was sent to customers. It warns of the potential for no anesthetic agent output. The letter contains short term recommendations for use of the device until a GE Healthcare representative contacts the facility to schedule a vist and implelment corrective action.
World wide-(including the States of IA, CA, MI, FL, MI, TX, IL, WI, OH, and countries of Finland, Singapore, Austria, South Africa, Sweden, Israel, Norway, Taiwan, Germany, Turkey, Portugal, Italy, Switzerland, Great Britain, and France
제품 설명
Fresh Gas Control (A-FGC1) Unit, a component of the S/5 Anesthesia Delivery Unit. The Anesthesia Delivery Unit is designed to mix and dose respiratory gases and to ventilate the patient. The delivery unit is the basis for a flexible anesthesia system with full monitoring and data management capabilities.