U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 The Metrix Company 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    74043
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2264-2016
  • 사례 시작날짜
    2016-06-23
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System/device, pharmacy compounding - Product Code NEP
  • 원인
    An issue was identified after a redesign of the auto compounding device. the compounder cannot meet the designed accuracy specification when dispensing into bags. after reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
  • 조치
    On 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form.

Device

  • 모델명 / 제조번호(시리얼번호)
    All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 66505 and 66510
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
  • 제품 설명
    Container fill and transfer sets, 250 mL and 500 mL, . || .
  • Manufacturer

Manufacturer

  • 제조사 주소
    The Metrix Company, 4400 Chavenelle Rd, Dubuque IA 52002-2655
  • 제조사 모회사 (2017)
  • Source
    USFDA