Biomet spine llc is recalling the set screw 5.5 mm rod due to the set screw being out of specification.
조치
Biomet Spine LLC (formerly Lanx) sent an Urgent Recall Notification letter to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions contact Lanx Customer Service.
Worldwide Distribution - US Nationwide in the states of: AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MO, NC, NE, NY, OH, OR, PA, TN, TX, VA, WI. and the countries of: Italy, and Mexico.
제품 설명
Set Screw 5.5 mm Rod - Model #7703-1600. || Product Usage: The titanium alloy set screw is used to seat a 5.5 mm titanium alloy, commercially pure titanium or cobalt chrome rod into a pedicle screw tulip in order to create a spinal fixation construct. The Set Screw is provided to the user non-sterile and is used across three Biomet Spine product lines; Silverton Spinal Fixation System (Silverton), Silverton-D Spinal Fixation System (Silverton-D) and Telluride Minimally Invasive Spinal Fixation System (Telluride). || The primary packaging for the Set Screws is an aluminum sterilization case that holds both implants and instruments used collectively as part of a system. Additionally, Set Screws can be packaged in a heat sealed polypropylene bag for single device distribution.
“The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.