Device Recall Siemens CentraLink Data Management System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Siemens Healthcare Diagnostics, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    69129
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2665-2014
  • 사례 시작날짜
    2014-08-19
  • 사례 출판 날짜
    2014-09-18
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-05-12
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel data.
  • 데이터 추가 비고
    unknown device name - Product Code N/A
  • 원인
    Siemens healthcare diagnostics has identified a task scheduler issue with the centralink" data management system, software version 14. the issue may cause the software to stop executing commands, including uploading validated results to the lis. the issue is related to an internal software timer that overflows after 24 days. automated rules triggered by task schedulers or by actions in the centralink user interface may also be affected. as each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.
  • 조치
    An Urgent Medical Device Correction (UMDC #10819364) letter was issued on August 19, 2014 to all affected customers. The letter instructed customers as follows: Siemens Healthcare Diagnostics is conducting a field correction for the CentraLink" Data Management System. To prevent the issue from occurring, log out of user sessions and restart the CentraLink services and translators as soon as it is convenient, then periodically restart going forward. Any restart frequency that occurs more often than every 24 days is acceptable and will prevent the issue from occurring. Refer to the Additional Information section for detailed instructions.

Device

  • 모델명 / 제조번호(시리얼번호)
    Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474
  • 유통
    Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Lebanon, Lithuania, Luxemburg, Malaysia, Mexico, Netherland, New Zealand Norway, Pakistan, Portugal, Peru, Qatar, Russia, Saudi Arabia, Singapore Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand United Arab Emirates, United Kingdom, Uruguay, Zambia.
  • 제품 설명
    Siemens CentraLink Data Management System, software version 14. || The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.