Material, embolization, neurovascular, polymerizing or precipitating - Product Code NPE
원인
Circadiance has determined that it is possible for certain smart monitor 2ps/psl monitors to exhibit intermittent operation of the nurse call feature. the firm has updated the device to increase the "nurse call" feature to ensure monitor alarms are transmitted to compatible nurse call systems.
조치
Circadiance mailed a Medical Device Removal SmartMonitor 2 PS/PSL, dated 04/28/2017, to affected customers to inform them of the issue, the actions to be taken and the actions the firm plans to take to correct the issue. Customers are told to test the functionality of the system prior to each use; and, complete the Business Reply Form. The form can be returned via 412-202-4583 or emailed to james.gianoutsos@circadiance.com. Customers are to contact Circadiance Customer Service at 724-858-2837 to receive an Return Material Authorization number; or, 1-888-825-9640 to repair their unit free of charge, or for questions or support.
Unique Device Identifier (UDI) 1030271, 1038140, 1038141, U1030271, and U1038140. All serial numbers of units with "REF", "MODEL" or standalone part number 1030271, 1038140, 1038141, U1030271 or U1038140 located on the products label are impacted by this notice. Monitors with the above part numbers that have a serial number starting with "CP" are NOT affected by this issue.
SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. || Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.