U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Medical computers and software - Product Code LNX
원인
Soft query results reporting versions 2.2.0.6.3 and 2.2.0.7.1 his may display incorrect collection data for a procedure or test result. clients using hbo star rr interface to query for results on a canceled test may see the previous occurrence of the test with the current date and time displayed as the collection date and time instead of the correct collection date and time sent by the scc interf.
조치
Client Notification Task Verbiage was sent through the Task Management System on December 3, 2003, identifying the software issue, the clinical impact, and corrective action. Customers were informed that the HIS may display incorrect collection data. Customers received the SQRR Service Pack with revised software, which will will not produce inaccurate information if there are no verified results. Customers were to acknowledge receipt of the task notification and indicate whether they would like the corrected software.
SCC Soft Computer proprietary Task Management System is used to manage client communication. Clients use this system to request services, receive and provide software updates for any issue.
Customers can contact SCC Soft Computer at 727-789-0100.
Nationwide Distribution: including states of GA, LA, and NH.
제품 설명
Soft Query Results Reporting Versions 2.2.0.6.3 and 2.2.0.7.1. || SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. || Soft Query Results Reporting version 2.2.0.6.3 was released for General Availability on 09/11/03. || Soft Query Results Reporting version 2.2.0.7.1 was released for General Availability on 01/08/04.