Laboratory information System - Product Code LNX
원인
Potential for patient data to be sent to the wrong patient record. a software change affected how the softlab handles recovery from certain types of table access conflicts. when such conflicts occur, this function has the potential to cause faulty writing to the database.
조치
Clients were notified through SCC Soft Computer's proprietary Task Management System On 03/22/201. All clients have received a hot fix that corrected the issue or made the hosparam change that prevents the issue.
SCC Soft Computer, Softlab Laboratory information System with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3.