U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Medical computers and software - Product Code LNX
원인
Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities. incomplete culture results may appear complete. scc soft computer is issuing a correction for softmic versions 4.0.4.7-4.5.1.3 software due to this potential hazard.
조치
SCC Soft Computer issued Risk-to-Health notifications to clients on July 19, 2010 via proprietary communication software. Corrective actions will be via programming correction via hot fix (software modification) or such so that Global variable is not used. Clients are asked to acknowledge receipt of this task and update the task with permission to load the required correction.
Worldwide Distribution: USA, including the states of CA, CT, DE, FL, IA, KY, MA, MA, MI, MN, NJ, NY, OH, PA, TX, VA, and WI, and the country of Canada.
제품 설명
SoftMic Versions 4.0.4.7, 4.5.1.3. || Version 4.0.4.7 Manufactured 02/03/2010. || Version 4.5.1.3 Manufactured 03/18/2010. || SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.