Laboratory Information System - Product Code JQP
원인
Modifications to diagnostic text may be: 1) saved to the database but not appear on the report sent to the physician; or 2) documented on the report, but not saved to the database.
조치
On 11/26/2014 SCC Soft Computer sent a Correction Communication to their consignees: Corrective Action: Versions 4.3.0.8 through 4.4.0.2 will have a mandatory hotfix created. Due to the amount of versions affected, the range of availability dates will be 11/24/2014 through 12/30/2014. Utility: A utility will be made available to be run on your Live system that will identify cases in which there is a disparity between the diagnosis text in the document and the database. The utility output will be provided for client review and validation.