Device Recall STAIA PACK hsE2 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Tosoh Bioscience Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79813
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2089-2018
  • 사례 시작날짜
    2018-03-05
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Radioimmunoassay, estradiol - Product Code CHP
  • 원인
    Asfotase alfa (strensiq) interferes with certain tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
  • 조치
    The firm TOSOH Bioscience Inc., sent an "URGENT: MEDICAL DEVICE RECALL" letter dated March 23, 2018 to affected customers. The letter described the product, problem and action to be taken. The customers were instructed to do the following: *Based on our revised labeling, the assays identified in Table 1 should not be used for patients who are being treated with Asfotase Alfa. " For patients that are being treated with Asfotase Alfa, use an alternative test method that does not utilize alkaline phosphatase technology. " Inform all medical professionals that assays utilizing alkaline phosphatase-based technologies must not be used for patients receiving Asfotase Alfa treatment. Continue to use the assays listed in Table 1 for any patients who are not receiving Asfotase Alfa treatment. " Complete and return the attached Acknowledgement Form to Tosoh Bioscience, Inc., within 15-days of receiving this notification via Fax: 1-650-636-8651, Email: biorecallresponse@tosoh.com or Regular mail: 3600 Gantz Rd., Grove City, OH 43123. " Maintain this notification with your laboratory records and forward this information to others who may have received this product. " If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the products identified in the table above, contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to call (800) 248-6764 or email: bernadette.oconnell@tosoh.com. Monday - Friday, from 9:00 AM to 5:00 PM (PST).

Device

  • 모델명 / 제조번호(시리얼번호)
    All Lot Codes
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.
  • 제품 설명
    ST-AIA PACK hsE2; Part Number: 025225 || Assay, Reproductive Hormone