The pre-vacuum steam sterilization (wrapped method) temperature in the instructions for use incorrectly listed 121-123¿c , instead of the correct 132-133¿c.
조치
Stryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
제품 설명
Stryker Irrigator Reusable Tips: || 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. || Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.