U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
The pre-vacuum steam sterilization (wrapped method) temperature in the instructions for use incorrectly listed 121-123¿c , instead of the correct 132-133¿c.
조치
Stryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
제품 설명
Stryker Irrigator Reusable Tips: || 5mm X 32cm Regular Tip, vented, Non-Reflective Finish. || Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.