Screw, fixation, bone - Product Code HWC
원인
Potential breakage -- product may break while inserting it into the bone during surgery. this may prolong surgery, or, if undetected during surgery, result in retained fragments that may cause metal sensitization or immune response, or necessitate further intervention to retrieve the titanium fragments.
조치
Stryker issued an Urgent Device Removal notification letter dated August 21, 2009 to consignees. Consignees were asked to acknowledge receipt and distribute the letter, track reponses, and destroy any returned material and document through weekly summary and records.
Stryker can be contacted at 408 754-2000 about this recall.