ventilator - Product Code CBK
원인
Some servo ventilators 300/300a and servo ventilator 900c/d/e should not be used with a system that may generate negative pressure below -100 cm h2o (closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
조치
Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Serial numbers 00410-24999. Ventilators in the specific range that are marked with one of the following labels are already equipped with the new type ot pressure transducer. 1. Upgrade EM008/02/1; 2. Upgrade SV300 66 82 745 SV900 66 82 747