Guide, surgical, instrument - Product Code FZX
원인
For certain lots, the incorrect raw material of annealed 17-4ph stainless steel was used instead of the correct material of heat-treated 304 stainless steel. this may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
조치
An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions:
Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product:
Option 1:
Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products.
Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: Synthes6734@stericycle.com
Option 2:
Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility.
Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.
Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). || The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.