System, blood culturing - Product Code MDB
원인
Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of streptococcus pneumoniae.
조치
The recalling firm made phone calls to their customers beginning 6/11/14 with follow-up letters dated 6/12/14 being issued via regular mail requesting the affected product be destroyed. Additional customers were later identified and additional phone calls were made on 6/16/14 and recall letters dated 6/16/14 were issued via regular mail.
Distribution was made to CO, IL, IN, NE, NJ, PA, SD, and a government facility located in VA.
제품 설명
Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.