Device Recall Transfer Spike (Blood processing accessory) 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Medtronic Cardiovascular Revascularization & Surgical Therap 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    51952
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1424-2009
  • 사례 시작날짜
    2009-02-25
  • 사례 출판 날짜
    2009-06-23
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2011-05-07
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Blood processing accessory - Product Code DWE
  • 원인
    Sterility may be compromised. non-vented caps that connect to the 3-way stopcock attached to the transfer spike could prevent ethylene oxide from contacting the inside surfaces of the stopcock during sterilization, leading to a potentially non-sterile product. use of the product may cause serious health consequences from potential sepsis/infection and the associated remote risk of death.
  • 조치
    Consignees were sent a Medtronic "Urgent Medical Device Recall Notice" letter dated March 5, 2009. The letter was addressed to Risk Manager. The letter described the product and problem and requested the consignee to complete, sign and return the certificate acknowledgement.

Device

Manufacturer

  • 제조사 주소
    Medtronic Cardiovascular Revascularization & Surgical Therap, 7611 Northland Dr N, Brooklyn Park MN 55428-1088
  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.