U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Ventilator, continuous, facility use - Product Code CBK
원인
It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.
조치
Philips Respironics has contacted all of its consignees by phone and instructed them to remove the potentially affected external battery packs from their inventory and any that may be on their patients/end users. These consignees have been instructed to quarantine batteries they have removed away from flammable materials. Product return/replacement instructions were provided.
For further information, contact Philips Respironics at 1-877-387-3311.
Worldwide Distribution -- United States (MT, IL, IA, NY, NC, TX, CA and FL), France, South Korea and Turkey.
제품 설명
Trilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. || Provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.