Wheelchair, powered - Product Code ITI
원인
Motion concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the vertical position system (trx vertical positioner).
조치
URGENT RECALL NOTIFICATION Letters (dated 8/26/2014) were sent to customers on 8/26/14 via Federal Express. Motion Concepts is requesting that all customers that have purchased a Motion Concepts TRx Vertical
Positioner system in between June 2011 - May 2014 to make arrangements for replacement of the knee support mount in accordance with the instructions enclosed in the letter. For questions, please contact us at 905-695-0134.