Device Recall TrypticaseTM Soy Agar with 5 Sheep Blood (TSA IITM), 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Becton Dickinson & Co. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77609
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0196-2018
  • 사례 시작날짜
    2017-05-22
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2018-02-26
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Culture media, antimicrobial susceptibility test, excluding mueller hinton agar - Product Code JSO
  • 원인
    Listeria monocytogenes contamination of non-sterile plated media.
  • 조치
    BD Life Sciences initiated communication with distributors by issuing Urgent Product Recall and Customer Response Form notices via UPS mailing on May 22, 2017. General customers were issued Urgent Product Recall and Customer Response Form letters via UPS mailing on May 25, 2017. - General customers were advised to visually inspect all non-sterile prepared media prior to use to assure that there is no surface and/or subsurface contamination. If the customer observes a plate without evidence of microbial contamination that grows Listeria, please consult the clinician to make sure this result is consistent with the patient's clinical presentation. - Distributors were advised to discontinue distribution of the lots and to discard any remaining inventory. Customers are advised to complete and return the enclosed response form to: Attention: Regulatory Compliance, Fax: 410-316-4258 or email to BDRC2@bd.com **Note to Distributors: Please provide BD Regulatory Compliance an Excel file or a list with name, address and telephone number of end-users who were shipped this product from their distribution branches so they can be notified. Mail or fax to: BD Regulatory Compliance 7 Loveton Circle Mail Stop 640, Sparks, MD 21152 Fax: 410-316-4258; Email: BDRC2@bd.com. Customers with questions please contact BD Support at 1-800-638-8663. For customers outside of the U.S., contact your local BD representative or distributor. For questions regarding this recall call 410-316-4000.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Number: 7032978 - Exp Date May 24, 2017  7048902 - Exp Date June 9, 2017
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, FA, HI, IA, IS, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT & WA; International - Canada, Guam and Japan
  • 제품 설명
    TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM), Catalog Number 221261 || Trypticase Soy Agar with 5% or 10% Sheep Blood is used for cultivating fastidious microorganisms and for the visualization of hemolytic reactions produced by many bacterial species.
  • Manufacturer

Manufacturer