Device Recall Varian 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Varian Medical Systems Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    25575
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0633-03
  • 사례 시작날짜
    2003-01-31
  • 사례 출판 날짜
    2003-03-12
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2003-09-17
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Accelerator, Linear, Medical - Product Code IYE
  • 원인
    The product has a software problem. this could result in a treatment without a motorized wedge where one had been planned.
  • 조치
    On 1/16/03 and 2/4/03 (2nd notification) the firm initiated the recall and their notification was via letters to inform its consignees about the new VARiS version 6.2.27 release.

Device

  • 모델명 / 제조번호(시리얼번호)
    Affected Serial Numbers: 44, 260, 315, 495, 746, 751, 873, 907, 984, 1098, 1131, and 9045
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Product was released for distribution to 7 hospitals in the US. Distribution was limited to the following States: CA, MA, NC, SC, PA, and LA. 1 (one) Canadian distribution includes Fraser Valley Cancer Center, Dept. of Radiation Oncology, 13750 96th Ave., Surrey, BC, V3V 1Z2, Canada. Foreign distribution includes Austria, Germany and Finland. The recall was appropriately extended to the user level; i.e., the hospitals/physicians who received the recalled product.
  • 제품 설명
    VARiS 1.4G and Vision/SomaVision Radiation Therapy Systems; VARiS versions 6.1.33 and 6.2.19 in conjunction with Eclipse 6.x/Soma Vision 6.x
  • Manufacturer

Manufacturer

  • 제조사 주소
    Varian Medical Systems Inc, 911 Hansen Way, Palo Alto CA 94303
  • Source