Phenotypic test kit, non-susceptible/elevated MIC organisms, cultured isolates - Product Code PTJ
원인
The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the vitek¿ 2 ast-gn colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by eucast/clsi).
조치
Customers were notified on or about 05/16/2017 and then notified again with an updated letter. Instructions to customers included requesting that the attached acknowledgement form be completed, distribute the recall notification to customers, and provide the notice to future customers.