Device Recall VITEK TWO ASTN085 REF 22 225 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Biomerieux Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    56365
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2295-2010
  • 사례 시작날짜
    2010-07-29
  • 사례 출판 날짜
    2010-08-27
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2010-11-02
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, test, automated, antimicrobial susceptibility, short incubation - Product Code LON
  • 원인
    Foil pouch was punctured by drug susceptibility cards during packaging, allowing moisture into pouch, resulting in inaccurate and/or incorrect drug susceptibility results.
  • 조치
    BioMerieux issued an Urgent Product Removal Notice, dated July 29, 2010 to consignees, identifying the reason for the recall and the affected lots. Customers were instructed to discontinue use of and destroy any remaining inventory of the affected lot, and to notify their customers who received the product. All consignees were instructed to complete and return the Customer Acknowledgement and Product Replacement Form. These cards will be used to distribute replacement cards. Consignees can contact BioMerieux at 1-314- 731-8667.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot number 345158910 Exp:2011/04/21
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution: Country of Canada only. No product was distributed in the US.
  • 제품 설명
    VITEK TWO AST-N085 REF 22 225 For In Vitro Diagnostic Use - each test card is packaged into a foil pouch with 20 pouches per box. The firm name on the label is bioMerieux, Inc., Durham, NC and bioMerieux, sa, France.This is an Export Only test kit that is not marketed in the U.S.
  • Manufacturer

Manufacturer