Device Recall VITROS 4600 Chemistry System Software 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Ortho-Clinical Diagnostics 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    61910
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2067-2012
  • 사례 시작날짜
    2012-05-21
  • 사례 출판 날짜
    2012-07-23
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2018-07-05
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • 원인
    Software automatically switches the lot when a vancomycin (vanc) or valproic acid (valp) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment.
  • 조치
    Ortho Clinical Diagnostics notified its foreign direct account consignees via email on May 21, 2012 and their direct account US consignees were sent an Urgent Distributor Safety Notice on May 22, 2012 via US Postal Service Priority Letter. The recall letter notified the consignees of the situation and that Ortho will replace their remaining inventory of non-expired GENs 14 or 15 for VITROS VALP Reagent only. Until then the consignees can continue using the VITROS VALP Reagent, GENs 14 or 15 until the replacement arrives. And to note that VITROS VANC Reagent, GENs 19 &20 are now expired. --- The letter also requested the consignees to complete and return the enclosed Confirmation of Receipt, no later than May 31, 2012. And to review the quality control beginning with the Calibration date of the VITROS VANC Reagent ( GEN 21 or above) or VITROS VALP (GEN & above) so that the QC results can be verified to be within acceptable ranges. --- For customers using VITROS VALP Reagent GEN 14 or 15, they were to respond as to the quantity of the non-expired GENs 14 and 15 remaining inventory; to load and use only one GEN of VITROS VALP Reagent at a time on the system until the replacement arrives. This notification letter should be posted by each VITROS System using the VITROS VALP Reagent or VITROS VANC Reagent or with the user documentation. And in addition, if these products were distributed then this information should be forwarded as well. For questions regarding this recall call 585-453-4110.

Device

  • 모델명 / 제조번호(시리얼번호)
    Version 2.0 & Below
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA (nationwide) and the countries of Australia, New Zealand, Brazil, Canada, Chile, Argentina, Paraquay, Ecuador, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, UK, Northern Europe, France, Germany, Italy, Spain and Portugal.
  • 제품 설명
    VITROS 4600 Chemistry System Software || For use in the in-vitro quantitative, semi quantitative, and qualitative measurement of variety of analytes of clinical interest.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Ortho-Clinical Diagnostics, 100 Indigo Creek Dr, Rochester NY 14626-5101
  • 제조사 모회사 (2017)
  • Source
    USFDA